Differences between the standard and sample
It is important that the reference material be of a similar reflectivity and have similar
properties to the sample. Otherwise large errors can be introduced, if for example a
poorly reflecting material is measured relative to a highly reflective PTFE reference
plate. The table below lists the appropriate reference materials that are recommended for
use.
Surface type Reflectivity Reference
Matt High PTFE reference plate
Matt Low
Labsphere diffuse
reflectance standards
Glossy High PTFE reference plate
Glossy Low NIST SRM 2021
The Troubleshooting section contains information on the available reference surfaces.
Figure 7: Highly diffuse samples are measured against a flat PTFE reference plate
Inaccuracy in the standard
If the standard material used to calculate a value is not accurate, this will hamper the
determination of the reflectance of the sample. Great care must be taken to keep reference
materials clean and unscratched.
For a detailed discussion on the factors affecting the measurement of diffuse reflectance
and transmission with integrating spheres, refer to reference (2).